melt flow indexer(MFI)for plastic flow rate meets the requirements of GB/ t3682-2000, and the main parameters of the melt indexer are also full
Full iso 1133-97, ASTM1238 standard requirements can be used for polyethylene, polypropylene, polystyrene, A
BS, polyamide, fiber resin, acrylic ester, polyformaldehyde, fluoroplastics, polycarbonate and other plastic materials
Melt mass flow rate (MFR) or melt volume flow rate (MVR) were measured.
Melt flow indexer technical parameters:
Temperature control range: room temperature ---450 ℃
Temperature control accuracy: ±0.2 ℃
Display resolution: 0.1 ℃
Zui large power consumption: ≤500 W
Temperature recovery time: 4 min
: piston position detection: (MVR)
Upper and lower ring distance: 30 mm
Displacement control accuracy: ± 0.1mm
Test range: 0.03-450g /10min
Standard component parameters:
Cylinder parameters: the inner hole Φ 9.55 ± 0.025 mm
Parameters: the piston is the head Φ = 9.475± 0.015 mm;
Piston head length H = 6.35±0.1mm
Mouth membrane parameters: extruding hole Φ = 2.095±0.005 mm
Mouth mould configuration: 2.095 ±0.005 / Φ Φ 1.18 ± 0.01 mm
Weight parameters:
Weight accuracy: ±0.5%
Basic configuration: A 0.325 kg (including pressure bar)
B 0.875 kg
C 0.960 kg
D 1.200 kg
E 1.640 kg
Note: weight can be added according to user's requirement