Melt Flow Index (MFI, MFR, MVR) Testing Equipment (also known as melt index instrument) is a special instrument for measuring the melt flow rate of thermoplastics under certain conditions.The melt flow rate (melt index) of thermoplastics refers to the mass or melting volume of the melt passing through the standard die capillary every 10 minutes at a certain temperature and load. It is expressed by MFR (MI) or MVR value, which can distinguish the viscous flow characteristics of thermoplastics in the molten state.It is of great significance to the quality assurance of raw materials and products of thermoplastics and chemical fiber.
Technical parameters:
temperature control
Range: 100 ~ 450℃
4h drift: no more than ±0.5 ℃
Temperature gradient: ≤1.0℃ (10 ~ 50mm at the upper end of the mouth mold inside the material barrel in the tropical region)
Resolution: 0.1℃
timing clock
Range: 0 ~ 6000s adjustable
Resolution: 0.1s ~ 1s
cutting device
Automatic cutting: set interval time (adjustable anywhere from 2 to 2000s), and automatically complete cutting
Inching cutting: manual button electric cutting
Manual cutting: manual rotary handle cutting
weight load
Accuracy: 0.5%
Range: 325 ~ 21600g discontinuous, the combined load can meet the standard requirements.
measuring range and accuracy
Determination range: 0.1 ~ 100g/10min
Where, when MFR is 0.04 ~ 30g/10min, it is suitable to measure flow velocity by mass method.When MFR> is 30g/10min, it is suitable to measure the flow rate by volume method.
National standard sample (PE) sample: repeat precision 2%
Accuracy:5%
host size (mm) : 460×320 & times;740
power supply: 220V AC 50Hz
weight: about 50kg (excluding weight of weights)